While the instrument is running, look through
10.
grating of the access door and verify that the
lamp is illuminated, then click
If the lamp is not illuminated, the replacement
halogen lamp may be defective. Replace the
lamp again. If the second lamp does not
illuminate, check the instrument fuses for failure
(see
page
122).
Perform the calibrations listed below after
11.
replacing the lamp. See:
•
Chapter 4, Performing the Regions of
Interest (ROI) Calibration
•
Chapter 5, Performing the Background
Calibration and Optical Calibration
•
Chapter 6, Performing the Pure Dye
Calibration
•
Chapter 7, Verifying the Instrument
Performance
Notes
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
.
Appendix A
Replacing the Halogen Lamp
Light should be visible
121