Pull the heated cover door to the front of the
9.
instrument, then close the access door.
Plug in, then power on the 7300/7500/7500 Fast
10.
system.
Create a new background plate (see
11.
"Creating a Background Plate"
To confirm that you have eliminated the
12.
contamination, perform a background calibration
run (see
"Performing the Background
Calibration and Optical Calibration" on
page
49).
If the contamination is still present, repeat
13.
steps 1 through
7, then go to
CHEMICAL HAZARD.
Ethanol is a flammable liquid and vapor.
Exposure causes eye, skin, and respiratory tract
irritation and may cause central nervous system
depression and liver damage. Read the MSDS,
and follow the handling instructions. Wear
appropriate protective eyewear, clothing, and
gloves.
Notes
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Appendix B,
on
page
125.)
step
14.
Appendix A
Decontaminating the Sample Block
Contamination
111
A