Creating A Plate Document For The Roi Calibration - Applied Biosystems 7300 Installation And Maintenance Manual

Real-time pcr system
Hide thumbs Also See for 7300:
Table of Contents

Advertisement

Creating a Plate Document for the ROI Calibration

Click
(or select File New).
1.
In the New Document wizard, click
2.
accept the default parameters.
It is not necessary to name or save the
Note:
plate document. The SDS software automatically
saves the ROI data to a set of calibration files on
the computer hard drive.
If a Network communications error is displayed,
click
. You can perform the ROI
calibration even if this error is displayed.
This message is displayed if your
Note:
computer is not connected to a network. If you
do not plan to connect this computer to a
network, select Do not show me this message
again, then select Never again on this machine.
If an ROI error is displayed, click
proceed with ROI calibration.
Notes
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Chapter 4 Performing the Regions of Interest (ROI) Calibration
Creating a Plate Document for the ROI Calibration
to
to
Select
Click
Click
Click
4
41