Chapter 5 Performing the Background Calibration and Optical Calibration

Creating a Plate Document for the Background Calibration

Creating a Plate Document for the Background Calibration
In the SDS software, click
1.
select File New).
Configure the New Document dialog box:
2.
Select Assay Background.
a.
Select Container 96-Well Clear.
b.
Select Template Blank Document.
c.
In the Operator field, enter your name.
d.
In the Comments field, enter any additional
e.
information that you want to save to the file
(such as the plate bar code).
In the Plate Name field, enter:
f.
Background_<date in DDMMYY format>
For example, the name for a plate run on
May 31, 2005 would be:
Background_310505.
Click
.
g.
In the SDS software, select File Save As.
3.
In the Save As dialog box:
4.
If the Save in field does not display SDS
a.
Documents, navigate to D: drive Applied
Biosystems SDS Documents.
Click
.
b.
Continue with
"Performing the Background
Calibration" on page
55.
Notes
54
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
(or
a
b
c
d
e
f
g