Chapter 6 Performing the Pure Dye Calibration

Creating a Plate Document for the Pure Dye Calibration

Creating a Plate Document for the Pure Dye Calibration
In the SDS software, click
1.
select File New).
Configure the New Document dialog box:
2.
Select Assay Pure Spectra.
a.
Select Container 96-Well Clear.
b.
Select Template Blank Document.
c.
In the Operator field, enter your name.
d.
In the Comments field, enter any
e.
information that you want to attach to the
file (such as the plate bar code).
Click
.
f.
It is not necessary to name or save
Note:
the pure dye plate document. The SDS
software automatically saves the pure dye
data to a calibration file on the computer
hard drive.
The Pure Spectra Calibration Manager is
displayed.
Continue with
"Preparing Plates and Performing the
Pure Dye Calibration" on page
Notes
72
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide