Chapter 3 Installing the Software and Powering On the Instrument
Troubleshooting
Troubleshooting – Function Tests (continued)
Condition: CCD, Filter Wheel, or Shutter Test Failures
1. Power off the instrument, wait 30 seconds, then power on the
instrument.
2. Perform the Instrument Function Test (see
3. If the test fails again, perform the ROI calibration as explained
in
"Performing the Regions of Interest (ROI) Calibration" on
page
35.
4. Determine if the ROI image appears in the ROI Inspector.
Yes – Continue with the installation.
No – Contact Applied Biosystems technical support (see
page
xiii) or your service representative.
Condition: Lamp Test Failure
1. Power off the instrument, wait 30 seconds, then power on the
instrument.
2. Perform the Instrument Function Test (see
3. If the test fails again, select Instrument
Status/Replacement.
4. If the Lamp Status/Replacement dialog box reports the lamp
status as Failed, replace the halogen lamp (see
Halogen Lamp" on page
5. Perform the Instrument Function Test (see
6. If the test fails again, contact Applied Biosystems technical
support (see
page
xiii) or your service representative.
Condition: Thermal Cycler Test Failure
1. Power off the instrument, wait 30 seconds, then power on the
instrument.
2. Perform the Instrument Function Test (see
3. If the test fails again, contact Applied Biosystems technical
support (see
page
xiii) or your service representative.
Notes
34
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
page
30).
page
30).
Lamp
"Replacing the
118).
page
30).
page
30).
Lamp
status

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

75007500 fast

Table of Contents