Perform the RNase P Experiment

Perform the RNase P Experiment

Materials
Required
When to Perform
the RNase P
Experiment
Purpose of the
Experiment
Notes
38
After you establish system communication, confirm the function of the system by
®
running a TaqMan
RNase P Fast Instrument Verification Plate.
®
• TaqMan
RNase P Fast Instrument Verification Plate
• Powder-free gloves
• Safety glasses
• Centrifuge with reaction plate adapter
Applied Biosystems recommends performing an RNase P experiment:
• After installing the system.
• After moving the instrument to another location.
• As needed to confirm the function of the instrument, depending on your laboratory
and regulatory requirements.
®
The TaqMan
RNase P Fast Instrument Verification Plate experiment verifies the
performance of the instrument. The RNase P plate is preloaded with the reagents necessary
for the detection and quantitation of genomic copies of the human RNase P gene (a single-
copy gene encoding the RNase moiety of the RNase P enzyme).
Each well contains:
®
• 1✕ TaqMan
Fast Universal PCR Master Mix, No AmpErase
• RNase P primers
• FAM
dye-labeled probe
• Known concentration of human genomic DNA template
The figure below illustrates the arrangement of the standard and unknown populations on
the RNase P plate. The RNase P plate contains five replicate groups of standards (1250,
2500, 5000, 10,000, and 20,000 copies), two unknown populations (5000 and 10,000
copies), and negative control wells (NC).
5K-copy
5K-copy
Unknown
Unknown
Population 1
Population 1
10K-copy
10K-copy
Unknown
Unknown
Population 2
Population 2
®
TaqMan
48-well RNase P Plate for
the StepOne
System
Applied Biosystems StepOne
®
UNG
5K-copy Unknown
Population 1
NC
STD 1250
STD 5000
STD 10000
10K-copy Unknown
Population 2
®
TaqMan
96-well RNase P Plate for
the StepOnePlus
System
and StepOnePlus
Real-Time PCR Systems
Installation, Networking, and Maintenance Guide
STD 2500
STD 20000