C-Link Protocol Commands

Hardware Configuration

Hardware
Configuration
B-34
Model 42i Trace Level Instruction Manual
meas mode
This command reports which measurement mode (NO/NO
or prereactor) is active. The example below reports that the measurement
mode is set to NO.
Send:
meas mode
Receive:
meas mode no
set meas mode mode
mode = | no/nox | no | nox | pre |
This command sets the instrument to NO/NO
mode, manual NO
mode, or manual prereactor mode. The example below
x
sets the instrument to the manual NO mode.
Send:
set meas mode no
Receive:
set meas mode no ok
pres comp
This command reports whether pressure compensation is on or off. The
example below shows that pressure compensation is on.
Send:
pres comp
Receive:
pres comp on
set pres comp onoff
These commands turn the pressure compensation on or off. The example
below turns pressure compensation off.
Send:
set pres comp off
Receive:
set pres comp off ok
temp comp
This command reports whether temperature compensation is on or off.
The example below shows the temperature compensation is off.
Send:
temp comp
Receive:
temp comp off
set temp comp onoff
These commands turn the temperature compensation on or off. The
example below turns temperature compensation off.
Send:
set temp comp off
Receive:
set temp comp off ok
contrast
This command reports the screen's level of contrast. The example below
shows the screen contrast is 50%, according to
, NO, NO
x
(auto) mode, manual NO
x
Table
B–7.