Operation
Service Menu

Calibrate Flow Zero

Calibrate Flow Span
3-74
Model 48i Trace Level-Enhanced Instruction Manual
when the instrument is in service mode. For more information on the
service mode, see
"Service
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Flow Calibration.
FLOW SENSOR CAL:
>ZERO
SPAN
SET DEFAULTS
RANGE
AVG
The Calibrate Flow Zero screen calibrates the flow sensor at zero flow.
Note The pump must be disconnected before performing the zero
calibration.
In the Main Menu, choose Service > Flow Calibration > Zero.
CALIBRATE FLOW ZERO:
CURRENTLY:
SET TO:
DISCONNECT PUMP AND
SAVE CURRENT FLOW
RANGE
AVG
The Calibrate Flow Span screen allows the user to view and set the flow
sensor calibrate span point.
Note An independent flow sensor is required to read the flow, then the
operator enters the flow value on this screen to perform the calibration.
In the Main Menu, choose Service > Flow Calibration > Span.
Mode" earlier in the chapter.
121
1.0000
DIAGS ALARM
0.000 LPM
0.000 LPM ?
DIAGS ALARM
Thermo Fisher Scientific