3. Press the Tools key and choose the Reports item by pressing the adjacent soft key.
The following is displayed on the 882's LCD.
EDID
Misc
4. Select HDCP to access the HDCP compliance test menu.
The following is displayed on the 882's LCD.
!CompRpt
5. Select EditPCP to define the capabilities of the HDCP device under test.
The following is displayed on the 882's LCD.
1.2
:CTS Ver.
Sink
:DUT Type
+All
:Test Num-
+
:Src Max KSV
The following table describes the test parameters and their settings (gray = N/A).
Parameter
CTS Version
DUT type
Test Number
Sink 1.1 Features Supported
Sink 1.1 Audio Supported
Sink Short Read Ri
Packets
HDCP
EditPCP
-
Explanation
Specifies whether you wish to test to the HDCP CTS 1.1
or HDCP CTS 1.2. Please note you will not be able to
select CTS 1.2 if the HDCP CTS 1.2 option is not
enabled on the 882EA.
Specifies the type of device under test. This can be one
of Sink, Source, Repeater3AB, or Repeater3C. For this
test, select Sink.
Specifies which test to run. Or enables you to select all
tests. Please note that Quantum Data highly recom-
mends running the tests individually.
Indicates whether the sink DUT supports
Advanced_Cipher mode and Enhanced Link Verifica-
tion. The values are + for yes and - for no.
Indicates whether the sink DUT supports audio output.
The values are + for yes and - for no.
Indicates whether the sink DUT will respond to short
reads of the Ri register. The values are + for yes and -
for no.