Chapter 11 Testing EDID for HDMI
Parameter
Dual-Link DVI
Lipsync Indicated
Dual Latencies
Setting up 882 for HDMI sink device EDID compliance
testing
Follow these procedures to set up the 882 for HDMI sink testing for EDID compliance. You
will use this procedure regardless of whether you are running the EDID compliance test
through the front panel or the Compliance Controller.
To set up the 882 for HDMI sink EDID compliance testing:
1. Connect an HDMI-to-HDMI cable between the HDMI OUT connector on the 882 and
the HDMI connector on the display under test.
2. Press the Source key and choose a valid HDMI format (for example, DMT0660) by
pressing the adjacent soft key.
Alternatively, to select the format through the command line interface, enter the
following command:
FMTL
DMT0660
FMTU
3. Activate the HDMI-H interface on the output port as follows:
Explanation
Indicates whether the sink under test supports dual-link
DVI. The values are + for yes and - for no.
Indicates whether the EDID of the sink under test indi-
cates lipsync latency values. The values are + for yes
and - for no.
Indicates whether the sink under test's audio and video
latency is substancially different when handling inter-
laced video formats versus when handling progressive
video formats. The values are + for yes and - for no.