882 Video Test Instrument User Guide (Rev A.35)
Note: You may have to select a valid HDMI format at this point.
a. Press the Interface key. A listing of signal interfaces appears on the 882's display
as shown below.
VGA
*
HDMI-D
HDMI-H
b. Choose the HDMI-H item by pressing the adjacent soft key. The interface is
activated and the port outputs the currently selected image and format.
VGA
HDMI-D
*HDMI-H
c. Alternatively, to activate the interface through the command line interface, enter the
following commands:
XVSI
4
// Selects the HDMI-H interface
ALLU
// Applies the interface setting to the 882
4. Press the Sink key, and then press the Options key. The following menu appears on
the 882's display:
-EDID Formats
+HP Bypass
-Emulate
5. Choose the CDF item by pressing the adjacent soft key. The CDF parameters appear
on the 882's display as shown below.
0:HDMI Output Count
1:HDMI Input Port
-CEC Root Device
>Primary AR: 4:3
6. Set characteristics of the display under test using CDF parameters. Scroll through the
list by pressing the + key.
Refer to
"Defining the capabilities of the display device under
definitions for each parameter in the CDF.
Note: If you are running the EDID compliance test through the Compliance Controller,
you do not need to configure the CDF through the front panel.
CVBS
S-VIDEO
SDI
CVBS
S-VIDEO
SDI
CDF
Cho
Cho.
test" on page 12 for
17