Quantum Data 881 User Manual page 851

Hide thumbs Also See for 881:
Table of Contents

Advertisement

PDAX:MXER
Class
Description
Command syntax
Query syntax
Query response
Example
Related commands
882 Video Test Instrument User Guide (Rev A.35)
Pixel Data Analyzer MaXimum number of ERrors
Signal analyzer - Pixel data
Determines the maximum number of detailed errors to record during the pixel analzyer
test.
PDAX:MXER records
records
number of errors to record for the pixel analysis test (0 - 1000)
PDAX:MXER?
records
PDAX:FRMS 100 //sets the number of frames to analyze in the pixel
data analysis test to 100
PDAX:MXER 1000 //sets the maximum number of errors to record in pixel
data analysis test to 1000
PDAX:RPTG 1
//enables the generation of reports for the pixel
data analysis test
PDAX:REFG 1
//enables auto reference capturing for the pixel
data analysis test
PDAX:CAPF
//captures a frame of pixel data for the pixel
data analysis test
PDAU:
//runs the pixel data analysis test
PDAX:FRMS, PDAX:RPTG, PDAX:REFG, PDAX:CAPF, PDAU