882 Video Test Instrument User Guide (Rev A.35)
1. For the 882E analyzer, connect the cables as follows:
a. Apply power to the Encryption Status Tester (EST) using the power adapter
supplied with the EST.
b. Connect an HDMI cable from the source device under test output to the EST HDMI
port labeled, "HDMI IN."
c. Connect an HDMI cable from the 882 HDMI Rx 1 port to the EST HDMI port labeled
"HDMI OUT."
d. Connect an HDMI cable from the 882 HDMI Tx 2 port to the EST HDMI port labeled
"HDMI."
e. Connect an HDMI cable from the 882 HDMI Tx 1 port to the HDMI input of the
display device (used to monitor the video).
The following diagram depicts the test setup for the 882EA:
2. Activate the HDMI-H interface on the output port as follows:
a. Press the Interface key. A listing of signal interfaces appears on the 882's display
as shown below.
VGA
*
HDMI-D
HDMI-H
b. Choose the HDMI-H item by pressing the adjacent soft key. The interface is
activated and the display is shown below.
VGA
HDMI-D
*HDMI-H
Alternatively, to activate the interface through the command line interface, enter the
following commands:
CVBS
S-VIDEO
SDI
CVBS
S-VIDEO
SDI