2. Open the XML file in a standard text editor.
A sample sequence XML file for the sequence named MySeq01 is shown below:
<?xml version="1.0" encoding="UTF-8" ?>
<DATAOBJ>
<HEADER TYPE="SEQ" VERSION="1.0" ></HEADER>
<DATA>
<STEP>
<FMT>/tffs0/Library/TestAnalog/DMT0660.xml</FMT>
<IMG>master.img</IMG>
<DELY>+3.0000000E+00</DELY>
</STEP>
<STEP>
<FMT>/tffs0/Library/TestAnalog/DMT0659.xml</FMT>
<IMG>SMPTEBar.img</IMG>
<DELY>+3.0000000E+00</DELY>
</STEP>
<STEP>
<FMT>/tffs0/Library/TestAnalog/DMT0675.xml</FMT>
<IMG>SMPTE133.img</IMG>
<DELY>+3.0000000E+00</DELY>
</STEP>
<STEP>
<FMT>/tffs0/Library/TestAnalog/DMT0685.xml</FMT>
<IMG>Flat.img</IMG>
<DELY>+3.0000000E+00</DELY>
</STEP>
<STEP>
<FMT>/tffs0/Library/TestAnalog/DMT0856.xml</FMT>
<IMG>Grill_11.img</IMG>
<DELY>+3.0000000E+00</DELY>
</STEP>
<STEP>
<FMT>/tffs0/Library/TestAnalog/DMT0680.xml</FMT>
<IMG>Hatch4x3.img</IMG>
<DELY>+3.0000000E+00</DELY>
</STEP>
<STEP>
<FMT>/tffs0/Library/TestAnalog/DMT0872.xml</FMT>
<IMG>ColorBar.img</IMG>
<DELY>+3.0000000E+00</DELY>
</STEP>
<STEP>
<FMT>/tffs0/Library/TestAnalog/DMT0885.xml</FMT>
<IMG>Ramp.img</IMG>
<DELY>+3.0000000E+00</DELY>
</STEP>
</DATA>
</DATAOBJ>
3. Make the edits in the XML file as necessary. An example is provided below showing
how to change the format, image, and delay for the second step in the test sequence.