Quantum Data 881 User Manual page 259

Hide thumbs Also See for 881:
Table of Contents

Advertisement

882 Video Test Instrument User Guide (Rev A.35)
To create and save a test sequence with HDMI digital formats using a text file:
1. Create HDMI digital versions of the formats you will be using in the test sequence.
We recommend using the Format Editor. Please refer to
using the Format
Editor" on page 186.
a. Load the format that you wish to use.
b. Specify the Color Encoding Method to set the digital video signal type as in the
example below:
c. Save the format under a different name. For example if you are using 1080p60, you
might name the new format as 1080p60h.
2. Using a text editor, open a text editor on your computer and enter the following
commands to create a three-step sequence named MySeq:
SEQN
// initializes the sequence edit buffer
SEQB
// begins a sequence editing session
STEP
1
// selects a step to be edited
SDLY
5.0
// sets the delay for the current step to five seconds
FMTL
1080p60h
// loads an HDMI digital format
IMGL
SMPTEbar
// loads an image
STEP
2
// selects a step to be edited
SDLY
5.0
// sets the delay for the current step to five seconds
FMTL
720p60h
// loads a second format (same image)
IMGL
Outline1
// loads an image
STEP
3
// selects a step to be edited
SDLY
5.0
// sets the delay for the current step to five seconds
FMTL
1080i29h
// loads a third format
"Creating custom formats
247

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

882

Table of Contents