PNAU
Class
Description
Command syntax
Example
Related commands
882 Video Test Instrument User Guide (Rev A.35)
Measure Pseudo Noise
Signal Analyzer - Pseudo-random noise
Loads the hardware required to generate the pseudo noise and measures all the related
required parameters.
PNAU
PNST 1 //Pseudo-random noise signal type set to QDI-BCM
PNSA 1 //Pseudo-random noise signal advance set to ACTIVE PIXELS
PNSM 0 //Pseudo-random noise signal mode set to AUTO
PNSP 0 //Pseudo-random noise signal procedure set to CONTINUOUS
PNAU
//Pseudo-random noise generated
GPER?
//Returns the pixel error rate (in errors per billion)
0.88
GNPT?
//Returns the number of pixels measured (in billions)
4.255
GCET?
//Returns the total number of errors detected on the red, green,
and blue components along with the total number of bad pixels
0,21,0,21
GFED?
//Returns data related to any error (first error only)
40018C,FFFFFF
GPER?, GNPT?, GCET?,
GFED?
485
Need help?
Do you have a question about the 881 and is the answer not in the manual?