Instrument Configuration

Contact Information

Thermo Fisher Scientific
operation upon exiting this screen. The following example shows the screen
when all analog outputs are set to "normal" operating mode.
In the Main Menu, choose Diagnostics > Test Analog Outputs > ALL,
Voltage Channel 1–6, or Current Channel 1–6.

SET ANALOG OUTPUTS:

SETTING:
OUTPUT SET TO:
 SET TO FULL SCALE
 SET TO ZERO
 RESET TO NORMAL
RANGE
AVG
The Instrument Configuration screen displays information on the
hardware configuration of the instrument.
Note If the analyzer is in service mode, pressing
selected item between yes and no. Zero/span valves needs to be set to ON
in order to do zero and span checks, and auto calibration needs to be set to
ON in order to do zero and span calibration resets.
In the Main Menu, choose Diagnostics > Instrument Configuration.
INSTRUMENT CONFIGURATION:
>I/O EXPANSION BOARD
ZERO/SPAN VALVES
PERM OVEN
LAG VOLUME
PERM DRYER
RANGE
AVG
CONVERTER
DILUTION
AUTO CALIBRATION
OUTPUT CONC CAL DATA
The Contact Information screen displays the customer service information.
In the Main Menu, choose Diagnostics > Contact Information.
ALL
NORMAL
DIAGS ALARM
YES
YES
YES
NO
NO
DIAGS ALARM
MOLY
NO
NO
NO
Model 42i Instruction Manual
Operation
Diagnostics Menu
will toggle the
3-65