Chapter 3
Meter Setup Menus
36
Orion Versa Star Pro Meter User Manual

Diagnostics Menu

Use the Diagnostics Menu to initiate a meter reset, test electrode stability, perform a meter self-
test or view meter and module serial numbers and software revisions.
Factory Reset – All meter settings are reset to factory defaults. Data log, calibration log and
methods are deleted. If System Access feature is enabled, all system restrictions (names,
passwords and access levels) are also deleted.
User Reset – All settings for the selected channel are reset to factory defaults. Data log,
calibration log and methods are retained.
Stability Test – Tests the stability of a pH electrode, ORP electrode or ion selective
electrode (ISE) by measuring the mV per minute drift and noise of the electrode.
Self-Test – Tests the keypad, meter accuracy and stirrer inputs. A self-test is also
performed automatically by the meter at each power up.
About Meter – Displays the meter and module software revisions and serial numbers.
Main Menu
Submenu
Factory Reset
Password
Channel 1
Channel 2
User Reset
Channel 3
Channel 4
Channel #
Stability Test
Channel #
Keypad check
Self Test
Accuracy check
Stirrer check
Meter
Channel 1
About Meter
Channel 2
Channel 3
Channel 4
Notes
Default password is 1111
The channel setup menu for the selected
channel will be reset to default settings.
Only channels with BNC connectors are
displayed in the stability test submenu.
The accuracy check is performed on
channels with BNC connectors only.
A stirrer probe (part # 096019) is
required to perform the stirrer check.
Serial No.
SW Rev.
######
#.##
The meter serial number and software
######
#.##
revision will be different from the module
######
#.##
serial numbers and software revisions.
######
#.##
######
#.##