Output Configuration Programming Example - Keithley 2500 User Manual

Photodiode meter
Hide thumbs Also See for 2500:
Table of Contents

Advertisement

12-10
Digital I/O Port, Output Enable, and Output Configuration
Output configuration programming example
Table 12-3
the Model 2500 as follows:
• Output enable: on
• Auto-off mode: on
NOTE
Table 12-3
Output configuration programming example
*RST
:FORM:ELEM CURR1
:SOUR1:VOLT 10
:OUTP:ENAB ON
:SOUR1:CLE:AUTO ON
:READ?
*Connect pins 8 and 9 of digital I/O port to simulate closed output enable switch.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
lists the command sequence for output configuration. These commands set up
Connect pins 8 and 9 of the digital I/O port together to simulate a closed output
enable switch. Otherwise, the unit will not turn on its output when the measure-
ment is made.
Command
Restore GPIB defaults.
Select channel 1 reading data.
Output 10V on channel 1.
Activate output enable.*
Enable auto-off mode.
Trigger and acquire readings.
Models 2500 and 2502 User's Manual