APPENDIX E
Because KSS removes operation 4 and part of operation 5 from the sweep cycle, it reduces the execution time of
commands involving a sweep cycle, tuning, or annotation. The following commands are some of those that
execute faster with KSS: TS, CR, El, RB, SF?
If a command does not require a sweep, or fetches information directly from the analyzer memory, its execution
time is not affected by KSS. This kind of command includes OL, RC, Rl, and KS94.
The following program illustrates how the KSS command saves time.
1 0
DIM
2 0
OUTPUT
4 0
FOR
OUTPUT 718;"CF4GZ;TS;El;MA;"
ENTER 718;A(N)
ENTER 718;B(N)
NEXT N
1 0 0
END
The program above stores the amplitude of the largest signal within a 1 MHz band centered about a frequency that
alternates between 4 and 6 GHz. This program takes approximately 40 seconds to run. The program takes only
25 seconds to run if line 20 is changed to read as follows:
OUTPUT 718;"IP KSS SPlMZ MS S2"
When the program has finished running, the CRT display is as shown in the following illustration.