The KS( command secures the contents of registers one through six. When the registers are secured, the SV and
SAVE commands cannot save more instrument states in the registers, but instead write "SAVE LOCK" on the
analyzer display. To save an instrument state in a locked register, first execute KS) to unlock the registers.
The recall function of the analyzer is not affected by this function.
OUTPUT 718;"KS(;"
The functions of the KS( command and the (,.,,, [,,, keys are identical.
The KS( command also protects the contents of any user-defined softkeys when the analyzer is under manual
operation. During manual operation, softkeys are loaded by pressing the
information erases the original contents of the softkey. If KS( has been execu ed, pressing
Lock Registers
Programming 195