Agilent Technologies 8566B Operating And Programming Manual page 231

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

FUNCDEF
Function Define
FUNCDEF
Item
COMMAND LIST
LENGTH
STRING
DELIMITER
The FUNCDEF command defines a program routine as a function label. After FUNCDEF is executed, the com-
mand list is executed whenever the function label is encountered.
Once the function label is defined, it can be loaded into a softkey which can be executed remotely, or locally from
the front panel.
When queried (?) , FUNCDEF returns the command list in an A-block data format.
(See KEYDEF and KEYEXC.)
122 Programming
string
I abet
del imi ter
Description/Default
Any spectrum analyzer commands from this
Remote section.
Two &bit bytes specifying length of command
list, in 8-bit bytes. The most significant byte is
first: MSB LSB
Must match. Marks beginning and end of
command list.
command
Range Restriction
Command list length is
limited to 2015 characters,
including carriage return (CR)
and line feed (LF).