Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

T3
External Trigger
The T3 command sets the analyzer to external trigger mode. This function triggers the analyzer sweep when an
external voltage passes through approximately 1.5 volts in a positive direction. The external trigger signal level
must be between 0 and 5 volts.
The functions of the T3 command and front panel (,,, trigger are identical. (See TM.)
OUTPUT 718; "T3;"
Programming 353