Agilent Technologies 8566B Operating And Programming Manual page 83

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

FULL SPAN INSTRUMENT PRESET (2 - 22 GHz)
Front Panel Preset
TRACE B, but not A or C.
To be precise:
FUNCTION:
DATA:
COUPLED FUNCTION:
TRACE:
MARKER:
INSTRUMENT STATE
SCALE:
REFERENCE LINE:
84 Manual Operation
Start Frequency
Stop Frequency
Reference Level
Hold
All set to a which corresponds to the following values:
Resolution Bandwidth
Video Bandwidth
Sweep Time
Attenuator
Center Frequency Step Size
A
B
A - B
off
States are saved including the current state. See @ a below.
Logarithmic, 10 dB/division
Display line off
Threshold off
INSTRUMENT STATE
2 GHz
22 GHz
0 dBm
3 MHz
500 msec full scale
10 dB, coupled to maintain < - 10 dBm
at input mixer
100 MHz entered in register
Clear-Write
Blanked, but information in memory
saved
off
5.5 divisions up
1 .O divisions up

Advertisement