Agilent Technologies 8566B Operating And Programming Manual page 313

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

KS92
Write in Display Units
DL
TH
Item
INTEGER
The KS92 command tells the analyzer to receive the display line (DL), threshold line (TH), marker normal (M2),
or delta marker (M3) position in display units. The program line below shows the KS92 syntax.
OWTJT 718 USING "K,B,K";"DL KS";92;"300HZ;"
The HZ keyword is this line functions as a terminator for the DL command. Without the HZ keyword, the DL
command will not work. When used with KS92, the DL, TH, M2, and M3 commands must be terminated with
one of the following keyword terminators: DM, DB, or HZ.
The KS92 command cannot be executed from the front panel.
This is the decimal ASCII equivalent and is transmitted to the analyzer as a single 8-bit byte.
l
204
Programming
KS
Description/Default
Represents display line, threshold, active
marker, or delta marker amplitude in display
units.
l
D M
H Z
Range Restriction
1 to 1022