Agilent Technologies 8566B Operating And Programming Manual page 258

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

Item
Carriage Return
Line Feed
etx
The KSJ command is a diagnostic aid used for servicing the spectrum analyzer.
The KSJ command allows the DACs on the Al6 Scan Generator and the A19 Digital-Analog Converter to be
manually controlled from the front panel.
The functions of the KSJ command and the (,,,, m keys are identical.
KSJ
Description/Default
Sets all DACs to the specified value.
Sets all DACs to the specified value.
Asserts end of text.
J
DAC Control
Range Restriction
ASCII code 13
ASCII code 10
P r o g r a m m i n g 1 4 9
KSJ

Advertisement

Table of Contents
loading