Agilent Technologies 8566B Operating And Programming Manual page 287

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

KSI
Transfer B to C
The KS1 command transfers trace B to trace C.
Note trace C is not a swept, active function. Therefore, transfer trace information to trace C as follows:
1. Select single sweep mode (S2).
2. Select desired analyzer settings.
3. Take one complete sweep (TS) .
4. Transfer data.
This procedure ensures that the current settings of the analyzer are reflected in the transferred data.
10
20
30
31
LOCAL718
40
E N D
When transferring trace data from one trace to another, only the trace information from
addresses is transferred out of the total 1024 available display memory addresses. Information in address 1024
and addresses 2026 through 2047 is not transferred. (Addresses 2026 through 2047 are usually used for custom
graphics.)
The functions of the KS1 and BTC commands are identical.
The functions of the KSI command and the m [I-,DIl keys are identical.
178 Programming