10
ASSIGN @Sa TO 718
20
OTJTPUT @Sa; "Yp,"
30
OUTPUT @Sa;"FAG.P)GZ;"
40
OUTPUT @Sa; "S2;TS;"
50
OUTPTJT @Sa; "HNLOCK;"
60
OUTPUT @Sa;"Sl;"
70
END
80
Line 20:
Presets the instrument.
Line 30:
Sets a start frequency of 6.0 GHz which automatically selects the second harmonic.
Line 40:
Sets the analyzer to single sweep mode and takes one complete sweep of the current display.
Line 50:
Locks the second harmonic of the local oscillator.
Line 60:
Resets the analyzer to continuous sweep mode.
Line 70:
Sets the frequency range of the second harmonic.
The functions of the HNLOCK and KSt commands are identical.
(Continued) HNLOCK
Programming 127