Agilent Technologies 8566B Operating And Programming Manual page 223

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

FOFFSET
Frequency Offset
Item
INTEGER
The FOFFSET command selects a value that offsets the frequency scale for all absolute frequency readouts, such
as center frequency. Relative values, like span, and delta marker, are not offset.
After execution, the FOFFSET command displays the frequency offset in the active function readout. The offset
value is always displayed beneath the CRT graticule line, as long as the offset is in effect.
The following program returns an offset value of 100 MHz to the controller and prints it on the controller screen.
10 OUTPUT7l8;"FOFFSET100MZ;FOFFSET?;"
20 ENTER718;N
30
PRINTN
40
END
When queried (?), FOFFSET returns the offset value as a real number, followed by carriage-return/line-feed
(ASCII codes 13, 10). The end-or-identify state (EOI) is asserted with the line feed.
114 Programming