Agilent Technologies 8566B Operating And Programming Manual page 260

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

The KSL command disables the noise level function which displays the RMS noise level at the marker. (See
MKNOISE or KSM.)
KSL does not blank the marker. Use MKOFF or Ml to blank the marker. (Because MKOFF and M2 remove the
marker from the screen, they also disable the noise level mode.)
10 OIJTRJT718;"MKF50MZ;"
20 OUTPUT718;"KSM;"
30 OUTPUT718;"KSL;"
40 OUTPUT 718;"Ml;"
50
END
Line 10:
Positions marker at 50 MHz.
Line 20:
Selects noise level mode.
Line 30:
Turns off noise level mode.
Line 40:
Blanks marker.
The functions of the KSL command and [,,,, a keys are identical.
L
KSL
Marker Noise Off
P r o g r a m m i n g 1 5 1