The BXC command exchanges traces C and B, point by point.
Note that trace C is not a swept, active function. Therefore, exchange traces C and B as follows:
1.
Select single sweep mode (SNGLS) .
2.
Select desired analyzer settings.
3.
Take one complete sweep (TS) .
4. Exchange data.
This procedure ensures that the current settings of the analyzer are reflected in the transferred data.
When transferring data from one trace to another, only amplitude information is exchanged, located in display
memory addresses 1025 through 2025 and 2049 through 3049.
The functions of the BXC and KSi commands are identical.
B X C
Exchange B and C
P r o g r a m m i n g 4 5