Agilent Technologies 8566B Operating And Programming Manual page 117

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

An HP-IB transmission may be disrupted if the analyzer LINE power is cycled. An
analyzer should be connected to an
Similar HP-IB disruption may result from pressing @ when the HP-IB is active.
Thus, a local lockout is recommended during HP 8566B automatic operation.
After analyzer power-on, wait 5 seconds before addressing it.
Shift Function Codes
Programming a shift function requires a code sequence similar to the manual procedure for activating a shift
function; that is, press (,,,,, , then press the key with the function's code (the front panel blue character).
For example, to select the video averaging shift function, blue code G, execute
About half of the shift key function codes require ASCII lower case letters or symbols. Spaces are not allowed
between the shift command, "KS" and the shift function, in this case "G".
CAUTION
operating
activates wdeo averaglng
HP-IB only with POWER ON.
P r o g r a m m i n g 7