Agilent Technologies 8566B Operating And Programming Manual page 64

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

SCALE
REFERENCE
AND
Modulated AM signal displayed in the 10 dB/division
scale shows the carrier, its sidebands, and distortion
products.
Linear scaling enables the observation of the sidebands proportional to the carrier.
LIN
Press (-J
As in the MARKER m example, Chapter 6, a
direct readout of the percent modulation can be
made.
The fractional readout is one -half the modulation
index (only one sideband is measured).
%AM = 2(.25) x 100 = 50%.
Note that the carrier signal need not be placed at the
reference level for an index ratio measurement.
Change to a logarithmic scale with m and change
the dB/division with @ @ .
The sidebands are 12 dB down from the carrier, veri-
fying the earlier measurement results.
LINE
LOG