FPKA
Fast Preselector Peak
The FPKA command automatically adjusts the preselector frequency to yield the greatest signal level at the active
marker. The FPKA command peaks the preselector faster than the preselector-peak command, PI?
Although this command can be executed in all frequency spans, it performs best when the instrument is in zero
span. Use the standard preselector peak for all other frequency spans.