Agilent Technologies 8566B Operating And Programming Manual page 279

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

Sample Detection
The KSe command selects the sample detection modeefor displaying trace information. The KSe command is
identical with manual operation of the front panel m I-] keys.
In sample mode, the instantaneous signal value of the final analog-to-digital conversion for the sample period is
stored in trace memory. As sweep time increases, many analog-to-digital conversions occur in each period, but
only the final signal value is stored and displayed.
Sample detection mode is automatically selected for video averaging and noise level measurements.
The above program line selects the sample detection mode of the analyzer.
170 Programming