Agilent Technologies 8566B Operating And Programming Manual page 379

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

Delta Marker
Item
REAL
The M3 command computes the frequency and amplitude difference of the active marker and a special marker,
called the delta or differential marker. These values are displayed in the display readout.
Differential value = active marker frequency - delta marker frequency
Differential value = active marker amplitude - delta marker amplitude
If a delta marker is not on screen, MKD places one at the specified frequency, or at the right side of the CRT If an
active marker is not on screen, MKD positions an active marker at center screen. (The active marker is the number
1 marker, unless otherwise specified with the MKACT command.)
270 Programming
DELTA MARKER
Description/Default
Selects delta marker frequency. Default value
for units is Hz.
Range Restriction

Advertisement

Table of Contents
loading

Table of Contents