C O U P L E D F U N C T I O N
When the center frequency is activated with step size in MANUAL, the active function readout includes both the
center frequency and the step size value.
DATA Entry
Example
Surveillance of a wide frequency span sometimes requires high resolution. One fast way to achieve this is to take
the span in sequential pieces using a tailored center frequency step. This example looks from 0 Hz to 2.5 GHz in
50 MHz spans.
First set the span and center frequency: For a span of
M H z , p r e s s (5) m m @
center frequency to 25 MHz with [z) @
Ii
Changes the step size in display unit increments.
Changes the step size in steps equal to one-tenth of the frequency span.
Selects a specific step size to a resolution equal to the current center fre-
quency readout.
S e t t h e
CENTER FREQUENCY STEP SIZE
I
I
I
I
Manual Operation 77