MKPX
Marker Peak Excursion
The MKPX command specifies the minimum signal excursion for the analyzer internal signal-identification routine.
The default value is 6 dB. In this case, any signal with an excursion of less than 6 dB on either side is not identified.
If MKPK HI (peak search) were executed on such a signal, the analyzer would not place a marker at the signal
peak.
OUTPUT 718;"MKPX 8dB;"
246 Programming