Line 130:
Places trace B in the view mode. This is necessary to prevent the analyzer from writing
over the data placed back into trace B by KS125.
Lines 40 to 150:
Sets the analyzer display address to 1024 with the DA command and sends the KS125
command to the analyzer. The "125" in KS125 is sent as a single binary byte.
Line 160:
Writes the integer string B-store, which contains the display memory values for the origi-
nal trace B display, into the analyzer trace B memory, restoring the original trace B display.
The KS125 command cannot be executed from the front panel.
This is the decimal ASCII equivalent and is transmitted to the analyzer as a single 8-bit byte.