The KS# command turns off the YIG#tuned mixed (YTX) self-heating correction factor. The functions of the KS#
command and the front panel (,., IL
keys are identical.
Normally, preselector tracking is controlled by the preselector digital-to-analog converter (DAC) , and the thermal
correction factor determined by the microprocessor. When KS# is executed, the thermal correction factor is not
applied to the preselector tracking. Therefore, a DAC value is the only condition affecting preselector tracking.
Executing an instrument preset (IP) is the only way to re-enable the YTX self-heating correction factor.