Agilent Technologies 8566B Operating And Programming Manual page 345

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

Marker Delta
Item
REAL
The MKD command computes the frequency and amplitude difference of the active marker and a special marker,
called the delta or differential marker. These values are displayed in the display readout.
Differential value = active marker frequency - delta marker frequency
Differential value = active marker amplitude - deIta marker amplitude
If a delta marker is not on screen, MKD places one at the specified frequency, or at the right side of the CRT If an
active marker is not on screen, MKD positions an active marker at center screen. (The active marker is the number
1 marker, unless otherwise specified with the MKACT command.)
The MKD command function is identical with that of the M3 command, and similar to that of the front panel m
key
236 Programming
MKD
Description/Default
Selects delta marker frequency. Default units is
HZ.
DELTA MARKER
real
Range Restriction