Item
INTEGER
LOCAL COMMAND
The KS127 command sends data, formatted in 2-byte binary, to the analyzer display memory All of the display
memory addresses can be written to with a single execution of KS127. The syntax of the KS127 command
requires a specified starting address that immediately precedes KS127. Specify the starting address with the DA
command.
If the controller is instructed to write to more addresses than there are between the specified starting address and
the last address in display memory, 4095, then a "wrap around" occurs, and the remaining display memory values
are sent to successive addresses starting at address p).
10
OPTION BASE 1
2 0
INTEGER B-&or@ 1001)
3 0
!
4 0
OUTPUT 718;"A4;Bl;TS;B3;"
50
OUTPUT 718 USING "K,B,#";"02;DA 1024;KS";123
60
ENTER 718 USING "VP";B-store( *)
70
OUTPUT 718;";Al;Bl;"
80
LOCAL 718
90
PRINT "CHANGE ANALYZER DISPLAYPRESS CONTINUE"
110 1
120 OUTPUT 718;"B3;"
130 OUTPUT 718;"DA 1024;"
140 OUTPUT 718 USING "K,B,#";"KS";l27
160 OUTPUT 718 USING "W,K";B-store( *);";'I
1 7 0
E N D
The KS127 command cannot be executed from the front panel.
This is the decimal ASCII equivalent and is transmitted to the analyzer as a single &bit byte.
l
l
KS
127
Description/Default
Represents amplitude data.
Return spectrum analyzer to local control.
Controller dependent.
KS127
Write to Display Memory
l o c a l
command
Range Restriction
Programming 2 11