SCALE AND REFERENCE LINE
The DISPLAY LINE can be positioned anywhere within the graticule. When activated after LINE power ON or
DISPLAY LINE m erases the line and readout from the CRT display but does not reset the last position. If the
display line is activated again before LINE power ON or [a;l:l , it will return to its last position.
DISPLAY LINE position is always accessible for HP-IB and TRACE m , even if never activated. See Chapter 5,
TRACE ARITHMETIC.
The DISPLAY LINE readout has the same number of significant digits as reference level.
DATA Entry
Example
When the amplitude of a number of signals in the same span require a quick readout, the DISPLAY LINE can be
used.
Activate the DISPLAY LINE with @
With
and read out its absolute amplitude level.
Moving the DISPLAY LINE to each signal reads out its amplitude.
Moves the line about one division for each full turn. The line moves in
display unit increments.
Moves the line one-tenth of the total amplitude scale per step.
Positions the line to the exact entry level. Entry may be in mV, FV -I- dBm,
a -94.9
DISPLAY LINE
Manual Operation 65