Agilent Technologies 8566B Operating And Programming Manual page 163

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

CNVLOSS
Conversion Loss
Item
INTEGER
The CNVLOSS command offsets the reference level. The reference level is calibrated when the analyzer is in
external mixing mode by compensating for the conversion loss of an external mixer with the CNVLOSS offset.
If a harmonic of the analyzer LO is not locked before the CNVLOSS command is executed, the message "NOT
HARMONIC LOCKED" is displayed on the CRT and the CNVLOSS command is ignored.
The analyzer stores the value of the offset entered with the CNVLOSS command. One offset value is stored for
each of the external mixing bands. The offset value remains as long as the band is locked. The offset value is
erased whenever the band is unlocked. (See HNUNLK or KSQ.)
5 4 P r o g r a m m i n g
Description/Default
Selects reference level offset to
compensate for conversion loss.
Range Restriction