Agilent Technologies 8566B Operating And Programming Manual page 303

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

KS=
Automatic Preselector Tracking
Use the KS= command to reinstate automatic preselector tracking after KS/ has been executed. Normally, the
center of the preselector filter automatically tracks signal responses in the four frequency bands of the 2 to 22 GHz
range.
The KS/ command allows manual adjustment of the preselector tracking. The PP command automatically opti-
mizes preselector tracking at any one frequency.
OUTPUT 718; "KS = ;"
The functions of the KS = command and the (,,lrrl & keys are identical.
(See KS/ and PI?)
194 Programming

Advertisement