Agilent Technologies 8566B Operating And Programming Manual page 353

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

MKPAUSE
Marker Pause
Item
REAL
The MKPAUSE command pauses the sweep at the active marker for the duration of the delay period.
OUTPUT 718;"MKPAUSE 100;"
When queried (?), MKPAUSE returns the value of the delay period as a real number followed by a
carriage-return/line-feed (ASCII codes 13, 10). The end-or-identify state (EOI) is asserted with line feed.
To turn pause off, turn off markers.
244 Programming
Description/Default
Delay time in seconds.
DELAY
Range Restriction
0 to 1000 seconds.