Agilent Technologies 8566B Operating And Programming Manual page 423

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

SMOOTH
Smooth
Item
TRACE LABEL
NUMBER OF
POINTS
The SMOOTH command smooths the trace according to the number of points
Increasing the number of points increases smoothing.
OUTPUT 718; "SMOOTH TRA 23; "
P r o g r a m m i n g
3 1 4
SMOOTH
Description/Default
Alpha character, User-defined label declared in
TRDEF statement.
Integer representing number of points for
running average.
n u m b e r o f p a i n t s
Range Restriction
AA-22 and _
1 < number of points < 39
Must be odd number.
specified for the running average.