Agilent Technologies 8566B Operating And Programming Manual page 87

Spectrum analyzer 100 hz - 2.5 ghz/2 - 22 ghz
Table of Contents

Advertisement

GENERAL DESCRIPTION
This chapter describes access and use of the B KEY FUNCTION.
Shift functions supplement a front panel function or provide unique measurement capabilities. The @ functions
are not named on the front panel but are coded by the blue characters beside the keys. For- example, the
frequency offset function is designated by the code V On the front panel the code V is found in the FUNCTION
section :
The shift functions are activated by pressing (WTII and then the front panel key with the appropriate blue code. A
complete summary of shift FUNCTION
shift functions at the end of this chapter.
Example
Activate the shift function V (frequency offset) with
shift light on
press
press
shift light off and offset function activated
The shift light can always be turned off with I--u] , which returns the front panel keys to their designated function.
DATA Entry
An active shift function value is read out and identified in the active function area of the display the same as any
other function using DATA entry Once the data has been entered, any other function can be activated. The shift
function will retain its last value until the a key is pressed, or the LINE switch is switched to STANDBY
Chapter 11
V
is in this chapter under FUNCTION SUMMARY There is an index to all
S
GENERAL
DESCRIPTION
Manual Operation 89

Advertisement

Table of Contents
loading