The KSt command (harmonic lock) limits the tuning range of the analyzer to a specific harmonic of the local
oscillator as selected by start and stop frequencies. The functions of the KSt command and the front panel (,.,
When executing a harmonic lock, use the following method to ensure the desired harmonic is selected:
1. Select desired harmonic by setting a start frequency within the desired band. (See Tuning Curves in Appendix
F for frequency range of each harmonic.)
2. Set analyzer to single sweep mode. Single sweep ensures the desired harmonic is selected when executing a
bank lock.
3. Execute KSt to lock the desired harmonic.
4. Set analyzer to continuous sweep mode.
5. Reset desired start and stop frequencies.
Remember, before you can change the frequency range to another harmonic, you must unlock the band with the
KSQ (harmonic unlock) command.
KSR (Diagnostics On) can be activated, at any time, to display the local oscillator harmonic number in the upper
left comer of the CRT display
10
ASSIGN @Sa TO 718
20
30
OUTPUT @Sa;"FA8.0GZ;"
40
80
80
70
OUTPUT @Sa;"FA8.0GZ;FB12.8GZ;"
80
E N D
Line 20:
Presets the instrument.
Line 30:
Sets a start frequency of 6.0 GHz which automatically selects the second harmonic.
Line 40:
Sets the analyzer to single sweep mode and takes one complete sweep of the current display
Line 50:
Locks the second harmonic of the local oscillator.
Line 60:
Resets the analyzer to continuous sweep mode.
Sets the frequency range of the second harmonic.
Line 70:
Harmonic Lock
(HNLOCK)
Programming 185